GLD140 Gold Performance Medical Switchers
140 Watt Multiple Output
FEATURES:
SPECIFICATIONS:
?
?
?
?
?
?
Compact size (4.5" x 8.5" x 1.95")
Power factor corrected to IEC 1000-3-2 Class A
Documented FMEA, WCA and EMC test results
EMI compliance to CISPR11, FCC Class B, IEC601-1-2
Approved to UL2601, IEC601-1 and
CSA 22.2 No. 601.1-M90 and EN60601: 1990
marked to LVD
Ac Input
90-264 Vac, 47-63 Hz single phase.
Output Power
Continuous output power 140 W with unrestricted convection
cooling, 180 W with 150 LFM of air.
Input Current
Active Power Factor Correction circuitry assures compliance
with IEC1000-3-2, Class A. Maximum input current at minimum
input voltage and full load is 3.0 A.
Efficiency
73 – 85 % at full rated load, nominal input voltage, depending on
model and load distribution.
Hold-up Time
26 ms total. 16 ms minimum from loss of ac input at full load until
activation of the power fail signal. Output voltages remain within
specified regulation limits for an additional 9 ms minimum after
power fail activates.
Overload Protection
Fully protected against short circuit and output overload. Short
circuit protection is cycling type power limit.
Output Noise
0.5% rms, 1% pk-pk, 20 MHz Bandwidth, differential mode.
Measured with noise probe directly across output terminals of
the power supply.
Transient Response
500 μs typical response time for return to within 0.5% of final
value for a 45% load step from any load greater than 10% of full
rated load, ? i/ ? t< 0.2 A μs. Maximum voltage deviation is 3.5%.
Remote Sense
Provided as a standard feature on V1 and V2.
Overvoltage Protection
Built in on V1, 2 and 3 outputs.
Voltage Adjustment
Factory set to specified voltage with user adjustable potentiom-
eter on V1, 2 and 3. User can adjust V1, 2 and 3 at least ±5% of
nominal output voltage.
Output Regulation
Regulation for all outputs is the maximum deviation from initial
set point under all line and load conditions. Initial set tolerance is
measured with all outputs at 50 % of full rated load.
No Load Turn-On/Standby
A minimum load of 3 A on V1 is required for proper regulation. If
not met, no degradation of reliability will occur.
Temperature Coefficient
0.03% / ° C typical on all outputs.
Input Protection
Internal ac fuses provided on both lines on all units.
Inrush Current
Thermal Shutdown
Provided as a standard feature. Designed to protect unit from
prolonged over temperature.
Power Fail
TTL / CMOS compatible output goes low (<0.5 V) 10 ms before
output voltage drops more than 4% below nominal voltage upon
loss of ac power.
Power Good
TTL / CMOS compatible output rises high 100 to 300 ms after V1
reaches regulation and should assure that sufficient energy is
stored in the input section to provide normal power fail /
shutdown.
Inhibit
Inhibit signal when pulled to the V1 output common will inhibit all
output voltages.
Fan Output
An additional thermally controlled 12 Vdc, 250 mA output suitable
for powering a dc fan is included in all models (Factory installed
cover with fan makes this output unavailable).
EMI/EMC Compliance
All models include built-in EMI filtering to meet the following
requirements:
EMC SPECIFICATION COMPLIANCE LEVEL
Conducted Emissions EN55011, Class B; FCC Class B
Static Discharge EN61000-4-2, 6 kV contact 8 kV air
RF Field Susceptibility EN61000-4-3, 3V/meter
Fast Transients / Bursts EN61000-4-4, 2 kV, 5 kHz
Surge Susceptibility EN61000-4-5, 1 kV diff., 2 kV com.
Conducted RF Susceptibility EN61000-4-6, 3V
Voltage Sags & Surges EN61000-4-11
Line Frequency Harmonics EN61000-3-2 Class A
EMC Compliance for Gold Series
Magnetic Emissions - Emissions will not exceed the limits of the
Army curve in MIL-STD-461D, RE101 when measured at 7 cm
from 30 Hz - 100 kHz over all nominal inputs and at full rated load.
Leakage Current
70 μA under normal conditions 120 Vac @ 60 Hz
Single fault conditions 280 μA, 264 Vac @ 50 Hz.
SAFETY AGENCY APPROVALS:
All models are Certified to be in compliance with the
applicable requirements of UL2601-1, CSA-C22.2 No. 601.1,
IEC 601-1/60601-1.
DESIGN VERIFICATION DOCUMENTS
The “ Gold ” series has undergone rigorous review and design
analysis. The following product documentation is available upon
request: 1. Failure Mode and Effects analysis (FMEA), 2. DVT
Data, 3. EMC / Susceptibility test results.
Inrush is limited by internal thermistors. The inrush at 240 Vac,
averaged over the first ac half-cycle under cold start conditions
will not exceed 37 A.
Condor D.C. Power Supplies, Inc., 2311 Statham Parkway, Oxnard, CA 93033
58
800-235-5929 ? 805-486-4565 ? FAX 805-487-8911 ? www.condorpower.com
相关PDF资料
GLD150-48G PWR SUPPLY MED SWITCH 150W 48V
GLDA01A1A SW ROTRY FIX LVR PLAS ROL SPDT
GLEA24C SWITCH TOP ROLLER PLUNGER DPDT
GLM110-524 POWER SUP SWITCHER 110W DUAL MED
GLM50D PWR SPLY MED SWITCHER 50W MULTI
GLM65-48 PWR SPLY MEDICAL 65W 48V
GLM75-24 PWR SPLY MED SWITCHER 75W 24V
GLM75IG PWR SPLY MED SWITCHER 75W MULTI
相关代理商/技术参数
GLD140B 制造商:SL Power Electronics 功能描述:- Bulk
GLD140C 功能描述:线性和开关式电源 140W MULTIPLE OUTPUT 5, 12, 15, -15V OUT RoHS:否 制造商:TDK-Lambda 产品:Switching Supplies 开放式框架/封闭式:Enclosed 输出功率额定值:800 W 输入电压:85 VAC to 265 VAC 输出端数量:1 输出电压(通道 1):20 V 输出电流(通道 1):40 A 商用/医用: 输出电压(通道 2): 输出电流(通道 2): 安装风格:Rack 长度: 宽度: 高度:
GLD140CG 功能描述:线性和开关式电源 140W MULTIPLE OUTPUT 5, 12, 15, -15V OUT RoHS:否 制造商:TDK-Lambda 产品:Switching Supplies 开放式框架/封闭式:Enclosed 输出功率额定值:800 W 输入电压:85 VAC to 265 VAC 输出端数量:1 输出电压(通道 1):20 V 输出电流(通道 1):40 A 商用/医用: 输出电压(通道 2): 输出电流(通道 2): 安装风格:Rack 长度: 宽度: 高度:
GLD140D 功能描述:线性和开关式电源 140W +5/+24/+12/-12V RoHS:否 制造商:TDK-Lambda 产品:Switching Supplies 开放式框架/封闭式:Enclosed 输出功率额定值:800 W 输入电压:85 VAC to 265 VAC 输出端数量:1 输出电压(通道 1):20 V 输出电流(通道 1):40 A 商用/医用: 输出电压(通道 2): 输出电流(通道 2): 安装风格:Rack 长度: 宽度: 高度:
GLD140D-CG 制造商:SL Power Electronics 功能描述:
GLD140DG 功能描述:线性和开关式电源 140W +5/+24/+12/-12V RoHS:否 制造商:TDK-Lambda 产品:Switching Supplies 开放式框架/封闭式:Enclosed 输出功率额定值:800 W 输入电压:85 VAC to 265 VAC 输出端数量:1 输出电压(通道 1):20 V 输出电流(通道 1):40 A 商用/医用: 输出电压(通道 2): 输出电流(通道 2): 安装风格:Rack 长度: 宽度: 高度:
GLD140EG 功能描述:线性和开关式电源 140W MULTIPLE OUTPUT 5, 24, 15, -15V OUT RoHS:否 制造商:TDK-Lambda 产品:Switching Supplies 开放式框架/封闭式:Enclosed 输出功率额定值:800 W 输入电压:85 VAC to 265 VAC 输出端数量:1 输出电压(通道 1):20 V 输出电流(通道 1):40 A 商用/医用: 输出电压(通道 2): 输出电流(通道 2): 安装风格:Rack 长度: 宽度: 高度:
GLD140F 功能描述:线性和开关式电源 140W 4 Output +5V/+15V/12V/-12V RoHS:否 制造商:TDK-Lambda 产品:Switching Supplies 开放式框架/封闭式:Enclosed 输出功率额定值:800 W 输入电压:85 VAC to 265 VAC 输出端数量:1 输出电压(通道 1):20 V 输出电流(通道 1):40 A 商用/医用: 输出电压(通道 2): 输出电流(通道 2): 安装风格:Rack 长度: 宽度: 高度: